The assessment of risks related to electromagnetic field (EMF) exposure in healthcare, required by Legislative Decree 81/08, must consider all working conditions. In MRI facilities, healthcare staff and patient companions may also be exposed to magnetic field gradients that extend beyond the gantry. This study mapped gradient exposure in areas typically occupied by operators and companions, using the fastest diagnostic sequences. The results were compared with the action levels of Legislative Decree 81/08 and the ICNIRP 2010 guidelines for both workers and the general public. Measurements on 1.5 T and 3 T MRI scanners allowed the identification of “safety distances” and provide useful guidance for EMF risk assessment in medical environments.

La valutazione dei rischi legati all’esposizione ai campi elettromagnetici (CEM) in ambito sanitario, prevista dal D.Lgs. 81/08 e successive modifiche, deve considerare tutte le condizioni di lavoro. Nelle strutture con sistemi di risonanza magnetica (RM), anche personale e accompagnatori possono essere esposti ai gradienti di campo magnetico che si estendono oltre il gantry. In questo studio è stata mappata l’esposizione ai gradienti nelle aree frequentate da operatori e accompagnatori, utilizzando le sequenze diagnostiche più rapide. I dati sono stati confrontati con i livelli di azione del D.Lgs. 81/08 e con le linee guida ICNIRP 2010 per lavoratori e popolazione generale. Le misure, condotte su scanner RM da 1,5 T e 3 T, hanno permesso di definire “distanze di sicurezza” utili alla valutazione del rischio CEM negli ambienti medici.

Characterization of the dispersion of magnetic field gradients inside MR installations for diagnostic imaging - Caratterizzazione della dispersione dei gradienti di campo magnetico all’interno delle installazioni di Risonanza Magnetica per la diagnostica per immagini

KIEFFER, CLAUDIA
2024/2025

Abstract

The assessment of risks related to electromagnetic field (EMF) exposure in healthcare, required by Legislative Decree 81/08, must consider all working conditions. In MRI facilities, healthcare staff and patient companions may also be exposed to magnetic field gradients that extend beyond the gantry. This study mapped gradient exposure in areas typically occupied by operators and companions, using the fastest diagnostic sequences. The results were compared with the action levels of Legislative Decree 81/08 and the ICNIRP 2010 guidelines for both workers and the general public. Measurements on 1.5 T and 3 T MRI scanners allowed the identification of “safety distances” and provide useful guidance for EMF risk assessment in medical environments.
2024
Characterization of the dispersion of magnetic field gradients inside MR installations for diagnostic imaging
La valutazione dei rischi legati all’esposizione ai campi elettromagnetici (CEM) in ambito sanitario, prevista dal D.Lgs. 81/08 e successive modifiche, deve considerare tutte le condizioni di lavoro. Nelle strutture con sistemi di risonanza magnetica (RM), anche personale e accompagnatori possono essere esposti ai gradienti di campo magnetico che si estendono oltre il gantry. In questo studio è stata mappata l’esposizione ai gradienti nelle aree frequentate da operatori e accompagnatori, utilizzando le sequenze diagnostiche più rapide. I dati sono stati confrontati con i livelli di azione del D.Lgs. 81/08 e con le linee guida ICNIRP 2010 per lavoratori e popolazione generale. Le misure, condotte su scanner RM da 1,5 T e 3 T, hanno permesso di definire “distanze di sicurezza” utili alla valutazione del rischio CEM negli ambienti medici.
File in questo prodotto:
File Dimensione Formato  
PRONTA_THESIS_SAN_MATTEO-46.pdf

accesso aperto

Dimensione 28.08 MB
Formato Adobe PDF
28.08 MB Adobe PDF Visualizza/Apri

È consentito all'utente scaricare e condividere i documenti disponibili a testo pieno in UNITESI UNIPV nel rispetto della licenza Creative Commons del tipo CC BY NC ND.
Per maggiori informazioni e per verifiche sull'eventuale disponibilità del file scrivere a: unitesi@unipv.it.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14239/30471